Specific Conductance Calibration Procedure
https://instrumentationtools.blogspot.com/2015/05/specific-conductance-calibration.html
Conductivity  is used to measure  the ability of an aqueous  solution  to carry an electrical  current. Specific  conductance  is the conductivity  value corrected  to 25 DC. 
![]()  | 
| Specific Conductance Calibration Procedure | 
Most  instruments  are calibrated  against  a single standard  which is near the specific  conductance of the environmental   samples.   The standard  can be either below  or above the specific conductance  of the environmental   samples.    A second standard  is used to check the linearity of the instrument  in the range of measurements. 
When  performing  specific conductance  measurement  on groundwater  or surface  water and the measurement   is outside the initial  calibration  range defined by the two standards,  the instrument will need to be re-calibrated  using the appropriate  standards . 
Specific  Conductance   Calibration  Procedure:
1. Allow  the calibration  standards  to equilibrate  to the ambient  temperature. 
2. Fill calibration  containers  with the standards  so each standard  will cover the probe and temperature  sensor.    Remove  probe  from its storage container,  rinse the probe with deionized  water or a small  amount  of the standard  (discard  the rinsate),  and'place  the probe into the standard. 
3. Select measurement  mode. Wait 'until the probe  temperature  has stabilized. 
4. Select calibration  mode, then specific  conductance.    Enter the specific conductance standard  value.   Make sure that the units on the standard  are the same as the units used by the instrument.   Ifnot,  convert the units on the standard  to the units used by the instrument. 
5. Select measurement   mode. The reading  should  remain  within manufacturer's specifications.                         If it does not, re-calibrate,   If readings  continue  to change  after re- calibration,  consult  manufacturer  or replace  calibration  solution. 
6. Remove  probe  from the standard,  rinse the probe with deionized  water or a small amount  of the second  standard  (discard  the rinsate),  and place the probe into the second  standard.   The second  standard  will serve to verify the linearity of the instrument.   Read  the specific conductance  value  from the instrument  and compare  the value  to the specific  conductance  on the standard.   The two values  should  agree within the specifications   of the instrument.   If they do not agree, re-calibrate.   If readings  do not compare,  then the second standard  may be outside  the linear range of the instrument.   Use a standard  that is closer to the first standard  and repeat  the verification.    If values  still do not compare,  try cleaning  the probe or consult  the manufacturer. 
7. After the calibration  has been completed,  rinse the probe with deionized  water  and store the probe  according  to manufacturer's   instructions. 
8. Record  the calibration  information  on calibration log sheet.
Note:   for projects  where  specific  conductance  is not a critical  measurement   it may be possible  to calibrate  with one standard  in the range of the expected  measurement.
