Oxidation-Reduction Potential (ORP) Sensor Calibration Procedure
https://instrumentationtools.blogspot.com/2015/05/ORP-Sensor-Calibration-Procedure.html
ORP stands for Oxidation-Reduction Potential. In practical terms, it is a measurement to oxidize contaminants. It's as simple as that
![]()  | 
| ORP sensor calibration | 
The Oxidation-Reduction Potential (ORP) Sensor measures the ability of a solution to act as an oxidizing or reducing agent. Use the ORP Sensor to measure the oxidizing ability of chlorine in swimming pools or to determine when the equivalence point has been reached in an oxidation-reduction reaction.
Checking ORP is a simple method to monitor the effectiveness of a sanitizer or the quantity of anti-oxidants in a liquid. In generalized terms for humans, a higher ORP is better for outside of the body, while a lower ORP is preferred for consumption due to the high anti-oxidant value.
The oxidation/reduction    potential  is the electrometric  difference  measured  in a solution  between an inert indicator  electrode  and a suitable reference  electrode.   The electrometric   difference  is measured  in millivolts  and is temperature  dependent. 
Calibration  or Verification  Procedure: 
1. Allow the calibration  standard  (a Zobell  solution:   read the warning  on the label before use) to equilibrate  to ambient  temperature. 
2. Remove  the probe  from its storage container  and place it into the standard. 
3. Select measurement  'mode. 
4. Wait for the probe temperature  to stabilize,  and then read the temperature. 
5. If the instrument  is to be calibrated,  do Steps 6 and 7.  If the instrument  calibration  is to be verified,  then go to Step 8. 
6. Look up the millivolt  (mv) value  at this temperature  from the millivolt  versus temperature  correction  table usually  found on the standard bottle  or on the standard instruction  sheet.  You may need to interpolate  millivolt  value between  temperatures. Select "calibration  mode", then  "ORP".  Enter the temperature-corrected    ORP value into the instrument. 
![]()  | 
| Oxidation-Reduction Potential (ORP) Sensor | 
7. Select measurement  mode.  The readings  should remain unchanged  within manufacturer's   specifications.    If they change,  re-calibrate.  If readings  continue  to change  after re-calibration,   try a new Zobell  solution or consult  manufacturer.   Go to Step 9. 
8. If the instrument  instruction  manual  states that the instrument  is factory calibrated,  then verify the factory calibration  against  the Zobellsolution.     If they do not agree within the specifications  of the instrument,  try a new Zobell solution.    If it does not agree, the instrument  will need to be re-calibrated  by the manufacturer. 
9. After the calibration  has been completed,  rinse the probe with deionized  water  and store the probe  according  to manufacturer's   instructions. 
10. Record  the calibration  information   on the calibration log sheet.

%2BSensor.png)