Temperature Sensor calibration procedure
https://instrumentationtools.blogspot.com/2015/05/temperature-sensor-calibration-procedure.html
Most  instrument  manuals  state there is no calibration  of the temperature  sensor, but the temperature  sensor must be checked  to determine  its accuracy.  This accuracy  check is performed at least once per year and the accuracy  check date/information   is kept with the instrument.   If the accuracy  check date/information   is not included  with the instrument  or the last check was over a year, the temperature  sensor accuracy  needs  to be checked  at the beginning  of the sampling  event. If the instrument  contains  multiple  temperature   sensors, each sensor must be checked.   This procedure  is not normally perform  in the field.  If the instrument  is obtained  from a rental company,  the rental company  should performed  the calibration  check  and include  with the instrument  documentation  that it was performed.
![]()  | 
| Temperature Sensor Calibration | 
Calibration Procedure:
1. Fill a container  with water  and adjust the water temperature  to below  the water body's temperature  to be measured.   Use ice or warm water to adjust  the temperature.
2. Place  a thermometer  that is traceable  to the National  Institute  of Standards  and Technology  (NIST) and the instrument's   temperature  sensor  into the water. Wait for both temperature  readings  to stabilize.
![]()  | 
| Temperature gauge calibration | 
3. Compare  the two measurements.    The instrument's  temperature   sensor  must agree with the reference  thermometer  measurement   within the accuracy  of the sensor (e.g.,±O.2°C).  If the measurements   do not agree, the instrument  may not be working properly  and the manufacturer   needs to be consulted.
4. Adjust  the water temperature  to a temperature  higher  than the water body to  be measured.
5. Compare  the two measurements.    The instrument's  temperature   sensor must  agree with the reference  thermometer   measurement   within  the accuracy  of the sensor  (e.g.,± 0.20 C).  If the measurements  do not agree, the instrument  may not be working properly  and the manufacturer  needs to be consulted.

