Our Site Moved to InstrumentationTools.com

Turbidity Sensor Calibration Procedure

The turbidity method is based upon a comparison of intensity of light scattered by a sample under defined conditions with the intensity of light scattered by a standard reference suspension. A turbidimeter is a nephelometer with a visible light source for illuminating the sample and one or more photo-electric detectors placed ninety degrees to the path of the light source. Note: the below calibration procedure is for a turbidimeter which the sample is placed into a cuvette. 

Some instruments will only accept one standard. For those instruments, the second, third, etc., standards will serve as check points. 

Calibration Procedure:

Turbidity sensor calibration procedure
Turbidity sensor calibration procedure
1. Allow the calibration standards to equilibrate at the ambient temperature. The use of commercially available polymer primary standards (AMCO-AEPA-l) is preferred; however, the standards can be prepared using Formazin (read the warning on the label before use) according to the EPA analytical Method 180.1. Other standards may be used if they can be shown that they are equivalent to the previously mentioned standards. 

2. If the standard cuvette is not sealed, rinse a cuvette with deionized water. Shake the cuvette to remove as much water as possible. Do not wipe dry the inside of the cuvette because lint from the wipe may remain in the cuvette. Add the standard to the cuvette. 

3. Before performing the calibration procedure, make sure the cuvettes are not scratched and the outside surfaces are dry and free from fingerprints and dust. If the cuvette is scratched or dirty, discard or clean the cuvette respectively. Note: some . manufacturers require the cuvette to be orientated in the instrument in a particular direction for accurate reading. 

4. Select a low value standard such as a zero or 0.02 NTU and calibrate according to manufacturer's instructions. Note: a zero standard (approximately 0 NTU) can be prepared by passing distilled water through a 0.45 micron pore size membrane filter. 

5. Select a high standard and calibrate according to manufacturer's instructions or verify the calibration if instrument will not accept a second standard. In verifying, the instrument should read the standard value to within the specifications of the instrument. If the instrument has range of scales, check each range that will be used during the sampling event with a standard that falls within that range. 

6. Record the calibration information on calibration log sheet.

Related

DP Level Transmitter with Zero Suppression Calibration

Schematic Diagram of a Open Tank DP Level Transmitter with Zero Suppression Calibration Open Tank DP Level Transmitter with Zero Suppression Calibration Formulas:  Min range =...

Open Tank DP Level Transmitter Calibration

Schematic Diagram of a open tank DP level transmitter installation Open Tank DP Level Transmitter Calibration Formulas:  Min range = SG x h + P0  Max range = SG x h + P 0 You c...

Oxidation-Reduction Potential (ORP) Sensor Calibration Procedure

ORP stands for Oxidation-Reduction Potential. In practical terms, it is a measurement to oxidize contaminants. It's as simple as that ORP sensor calibration The Oxidation-Reduction Potential (O...

Post a Comment Default Comments

emo-but-icon
:noprob:
:smile:
:shy:
:trope:
:sneered:
:happy:
:escort:
:rapt:
:love:
:heart:
:angry:
:hate:
:sad:
:sigh:
:disappointed:
:cry:
:fear:
:surprise:
:unbelieve:
:shit:
:like:
:dislike:
:clap:
:cuff:
:fist:
:ok:
:file:
:link:
:place:
:contact:

PopularRecentComments

Recent

Free PLC Training Courses and Tutorials

Free PLC training courses and articles to learn Programmable Logic Controllers with example ladder logic programs.Free PLC Training CoursesFree Siemens PLC CourseFree Wincc SCADA CourseFree Allen Brad...

PLC Study Material

PLC Study Material, PLC Theory, PLC Logic, PLC Interview Questions and Answers, PLC Ladder Logic, PLC Example Programs, Programmable Logic Controllers Study Materials. PLC S...

Magnetic Flow Meters Working Animation

Magnetic Flow Meters Working Animation Simply, Magnetic Flow Meters works on Faraday's Law of ElectroMagnetic Induction. The Two Green dots in the animation indicates electrodes. The Copper co...

Electromagnetic Flow Meters Working Principle

Magnetic flow meter, simply known as mag flow meter is a volumetric flow meter which is ideally used for wastewater applications and other applications that experience low pressure drop and with app...

Control Valve Selection Guide

Control valves play a major role in the everyday effort to increase process plant profitability and conserve energy. Proper selection of these valves can have a significant financial impact on the o...

Comments

JOHN MULINDI:

This is a good explanation that is very helpful to students aspiring to pursue instrumentation at university or collage level.

بلال:

Thank you very much for the information

Anonymous:

is there any laser sensor[transmitter or switch ] and is it suitable for industrial use , if it is suitable what is the advantage and disadvantages comparing to ultrasonic or radar , thanks a lot for ...

Anonymous:

for controlling [start/stop] of pump which is the suitable ,radar or ultrasonic?
I am planning to use non contact switch

Instrumentation Android App

item